Skip to main content

Table 2 Accuracy results

From: Development of a time-resolved fluorescence immunoassay kit for detecting canine coronavirus and parvovirus through double labeling

 

Spiked concentration

Mean ± SD

CV (%)

Recovery (%)

CCV

10

15.04 ± 1.09

7.23

108.06

(n = 4)

100

106.94 ± 1.65

1.54

102.71

 

500

511.50 ± 4.69

0.92

101.45

CPV-2

10

15.26 ± 0.92

6.01

110.28

(n = 4)

100

108.49 ± 1.89

1.74

104.26

 

500

510.26 ± 2.35

0.46

101.21