Table 6 PCR positivity comparison between selected RNA extraction automats using 20 patient samples

From: Flexible upscaling of laboratory PCR testing capacity at the Robert Koch Institute during the SARS-CoV-2 pandemic

Performance parameter

Q-Con

Q-HT

Q-Sym

Number of detected samples

20/20

18/20

18/20

 E-Gene and orf1ab detected

16/20

15/20

14/20

 Only E-Gene detected

3/20

0/20

3/20

 Only orf1ab detected

1/20

3/20

1/20

E-Gene mean Ct value

30.76

28.92

30.56

E-Gene range of Ct values

18.46–38.72

18.21–40.08

19.20–38.20

orf1ab mean Ct value

30.80

30.95

30.47

orf1ab range of Ct values

19.31–37.42

19.10–38.46

19.94–37.10

  1. PCR positivity was compared between three selected RNA extraction automats for detection of SARS-CoV-2 in 20 patient samples. PCR analysis was performed in duplicate by using the E-Gene/KoMa and orf1ab/c-myc duplex PCR assays.
  2. Q-Con: QIAcube Connect, Q-HT: QIAcube HT, Q-Sym: QIAsymphony