Skip to main content
Fig. 1 | Virology Journal

Fig. 1

From: Flexible upscaling of laboratory PCR testing capacity at the Robert Koch Institute during the SARS-CoV-2 pandemic

Fig. 1

Ct values for detection of SARS-CoV-2 in serial diluted cell culture supernatant were compared between the five tested automated RNA extraction methods. a Ct values for each tested sample obtained with the E-Gene rRT-PCR and b orf1ab rRT-PCT. c Difference of the Ct value obtained for a sample extracted by the listed automat to the mean Ct value measured for the same sample after extraction by the QIAcube Connect in E-Gene rRT-PCR, d orf1ab rRT-PCT, e E-Gene rRT-PCR after adjustment for the ratio of sample volume to RNA volume and f orf1ab rRT-PCT after adjustment for the ratio of sample volume to RNA volume. Q-Con: QIAcube Connect, Q-HT: QIAcube HT, Q-Sym: QIAsymphony, KFF: King Fisher Flex, MP96: MagNa Pure 96

Back to article page