Table 4 PCR efficiencies of various laboratories using ORF1ab and N Ct values from EQA samples

From: Assessing the comparability of cycle threshold values derived from five external quality assessment rounds for omicron nucleic acid testing

Lab code

ORF1ab

N

1/slope

intercept

r2

E (%)

1/slope

intercept

r2

E (%)

Tianlong nucleic extraction kit & BioGerm RT-PCR kit & Gentier PCR instrument

1110084

− 0.2827

43.86

0.9988

91.73

− 0.2737

45.56

0.9758

87.80

1110100

− 0.301

42.88

0.9796

99.99

− 0.301

44.72

1

99.99

1110155

− 0.1605

50.84

0.9995

44.71

− 0.301

43.64

1

99.99

110002B6

− 0.3345

42.39

0.9749

116.02

− 0.43

41.38

0.9423

169.15

Tianlong nucleic extraction kit & BioGerm RT-PCR kit & SLAN PCR instrument

1110051

− 0.2573

44.1

0.993

80.84

− 0.3627

41.83

0.9862

130.52

1110091

− 0.2509

44.89

0.9908

78.20

− 0.3763

41.77

1

137.85

Tianlong nucleic extraction kit & BioGerm RT-PCR kit & ABI7500 PCR instrument

1110006

− 0.301

42.05

0.9796

99.99

− 0.2676

44.71

0.9643

85.18

1110017

− 0.2737

43.36

0.9758

87.80

− 0.3345

43.57

0.9643

116.02

1110081

− 0.3763

40.04

0.9796

137.85

− 0.301

43.37

1

99.99

1110143

− 0.2573

43.99

0.9426

80.84

− 0.301

43.97

1

99.99

1110739

− 0.2509

44.89

0.9908

78.20

− 0.5017

40.65

0.9643

217.47

  1. Lab, laboratory; r2, the coefficient of determination; E, efficiency