Skip to main content

Table 4 Repeatability experiment of ILTV antibody detection

From: Development of a sensitive and specific xMAP assay for detection of antibodies against infectious laryngotracheitis and bronchitis viruses

ILTV detection

Samples

MFI

Average

Standard deviation

Intra Cv

Repeat Experiment 1

repeat 1

1145.5

1208.33

61.5433

5.09%

repeat 2

1268.5

repeat 3

1211

Negative

111.33

/

/

/

Repeat Experiment 2

repeat 1

1127.5

1143.00

55.8860

4.89%

repeat 2

1205

repeat 3

1096.5

Negative

84.33

/

/

/

Repeat Experiment 3

repeat 1

1046

1028.00

16.3707

1.59%

repeat 2

1024

repeat 3

1014

Negative

75.33

/

/

/

Inter MFI Average

1126.44

   

Standard deviation

91.30

   

Inter Cv

8.11%