Skip to main content

Table 10 Performance Summary of the Aptima, Roche and Abbott Assays

From: Analytical characteristics and comparative evaluation of Aptima HCV quant Dx assay with the Abbott RealTime HCV assay and Roche COBAS AmpliPrep/COBAS TaqMan HCV quantitative test v2.0

Assay

95% Limit of Detection WHO standarda (IU/mL)

Lower limit of Quantitation (IU/mL)

Linear Range (IU/mL)

Precision (log SD)

Aptima HCV Quant Dx

plasma 4.3 serum 3.9

10

10-100,000,000

0.14 at 2.06 log IU/mL (115 IU/mL)

Abbott RealTime

plasma 10.5 serum 7.2

12

10-100,000,000

0.10 at 1.96 log IU/mL (91 IU/mL)

Roche Ampliprep/ COBAS Taqman HCV test v2.0

plasma 11 serum 12

15

10-100,000,000

</= 0.09 at 2.48 log IU/mL (300 IU/mL)

  1. aSecond standard NIBSC 96/798 used for Aptima and Abbott. The third Standard NIBSC 06/100 for Roche Data taken from Aptima (11), Roche [14] and Abbott [15] package inserts