Skip to main content

Table 1 Recovery of AcroMetrix subtype B Linearity Panel in Aptima HIV Quant Dx

From: Analytical characteristics and comparative evaluation of Aptima HIV-1 Quant Dx assay with Ampliprep/COBAS TaqMan HIV-1 test v2.0

Target concentration (log copies/mL)

Average Aptima (log copies/mL)

Average Aptima-Target

Not detected

Not detected

Not detected

2.00

2.05

0.05

2.70

2.60

−0.10

3.70

3.65

−0.05

4.70

4.86

0.16

5.70

5.93

0.23

6.70

6.92

0.22