Skip to main content

Table 2 Detection of anti-HAV IgM in 10-fold serial dilutions of serum using ICA (SD Rapid test) and MEIA (AxSYM HAVAB-M 2.0)

From: Evaluation of an immunochromatographic assay for the detection of anti-hepatitis A virus IgM

Genotype

ICA developed in this study

MEIA (COI)

 

10-2

10-3

10-4

10-5

10-2

10-3

10-4

10-5

IA-1

++

++

++

-

+

(5.21)

+

(1.55)

-

(0.46)

-

(0.37)

IA-2

++

++

++

-

+

(5.14)

+

(1.78)

-

(0.49)

-

(0.36)

IIIA-1

++

++

+

-

+

(5.95)

+

(1.76)

-

(0.48)

-

(0.43)

IIIA-2

++

++

+

-

+

(4.72)

+

(1.35)

-

(0.56)

-

(0.31)

IB-1

++

++

+

-

+

(4.35)

+

(1.25)

-

(0.37)

-

(0.29)

IB-2

++

++

+

-

+

(4.58)

+

(1.01)

-

(0.38)

-

(0.29)

  1. (ICA: ++ strong positive, + weak positive, - negative)
  2. (MEIA: + positive, - negative, COI negative > 1 > positive)