Skip to main content

Table 1 Intra- and inter-assay variability of the established FQ-PCR assay

From: Development of TaqMan® MGB fluorescent real-time PCR assay for the detection of anatid herpesvirus 1

Dilution of standard (copies/reaction)

Intra-assay

  

Inter-assay

  
 

Mean

SD

CV (%)

Mean

SD

CV (%)

1010

7.0

0.30

4.3

6.9

0.34

4.9

109

9.9

0.19

1.8

10.1

0.23

2.3

108

13.4

0.25

1.9

13.6

0.34

2.5

107

16.9

0.29

1.7

17.0

0.36

2.1

106

20.8

0.33

1.6

20.9

0.40

1.9

105

24.7

0.37

1.5

24.7

0.49

2.0

104

28.6

0.49

1.7

28.7

0.77

2.7

103

32.3

0.52

1.6

32.2

0.90

2.8

102

35.4

0.81

2.3

35.3

1.16

3.3

101

37.2

2.31

6.2

37.5

2.51

6.7