Skip to main content

Table 3 Limit of detection for the S and C regions using the duplex real-time PCR assay

From: Real-time PCR assays for hepatitis B virus DNA quantification may require two different targets

HBV load(IU/ml)

S region

C region

Duplex

Positive results/total tests

Positive rate (%)

Positive results/total tests

Positive rate (%)

Positive results/total tests

Positive rate (%)

100

20/20

100

20/20

100

20/20

100

50

20/20

100

20/20

100

20/20

100

20

20/20

100

20/20

100

20/20

100

10

14/20

70

13/20

65

18/20

90

5

6/20

30

9/20

45

10/20

50

2.5

4/20

20

5/20

25

7/20

40