From: Performance of BioFire array or QuickVue influenza A + B test versus a validation qPCR assay for detection of influenza A during a volunteer A/California/2009/H1N1 challenge study
Diagnostic test
Sensitivity
Specificity
TP/(TP + FN)
%
95% CI
TN/(TN + FP)
BioFire
98/136
72.1
63.7–79.4
201/215
93.5
89.3–96.4
Rapid test
15/176
8.5
4.8–13.7
122/123
99.2
95.6–100