Skip to main content

Table 1 BioFire versus qPCR and rapid test versus qPCR outcomes

From: Performance of BioFire array or QuickVue influenza A + B test versus a validation qPCR assay for detection of influenza A during a volunteer A/California/2009/H1N1 challenge study

  qPCR positive qPCR negative
BioFire (n = 351)   
 Positive 98 14
 Negative 38 201
Rapid test (n = 299)   
 Positive 15 1
 Negative 161 122
  1. Results for matched tests performed by BioFire and qPCR or rapid test and qPCR