Skip to main content

Table 1 Intra and inter laboratory reproducibility

From: Validation of intra- and inter-laboratory reproducibility of the Xpert HPV assay according to the international guidelines for cervical cancer screening

Intra-laboratory reproducibilitya

 

Xpert HPV – AML Run 2

Total

Xpert HPV –AML Run 1

Negative

Positive

 

Negative

353

4

357

Positive

12

141

153

Total

365

145

510

Inter-laboratory agreementb

 

Xpert HPV UZ Ghent

Total

Xpert HPV –AML Run 1

Negative

Positive

 

Negative

356

1

357

Positive

10

143

153

Total

366

144

510

Inter-laboratory agreementc

 

Xpert HPV UZ Ghent

Total

Xpert HPV –AML Run 2

Negative

Positive

 

Negative

154

3

357

Positive

4

149

153

Total

157

153

510

  1. aThe overall HPV test agreement 96.9% (95% CI, 95.0–98.2%) KAPPA value of 0.925 (95% CI 0.888–0.961)
  2. bThe overall HPV test agreement 97.8% (95% CI, 96.2–98.9%) KAPPA value of 0.948 (95% CI 0.917–0.978)
  3. cThe overall HPV test agreement 98.6% (95% CI, 97.6–99.6%) KAPPA value of 0.967 (95% CI 0.943–0.991)