Skip to main content

Table 5 Repeatability experiment of IBV antibody detection

From: Development of a sensitive and specific xMAP assay for detection of antibodies against infectious laryngotracheitis and bronchitis viruses

IBV

Samples

MFI

Average

Standard deviation

Intra Cv

Repeat Experiment 1

repeat 1

2096.5

2177.67

75.2103

3.45%

repeat 2

2245

repeat 3

2191.5

Negative

177

/

/

/

Repeat Experiment 2

repeat 1

1770

1884.50

151.6583

8.05%

repeat 2

2056.5

repeat 3

1827

Negative

141.33

/

/

/

Repeat Experiment 3

repeat 1

1842

1832.00

106.3532

5.81%

repeat 2

1721

repeat 3

1933

Negative

127

/

/

/

Inter MFI Average

1964.72

   

Standard deviation

186.27

   

Inter Cv

9.48%