Skip to main content

Table 4 Intra- and inter-laboratory agreement for possible positivity outcome with REALQUALITY RQ-HPV Screen

From: Clinical validation of REALQUALITY RQ-HPV Screen according to the international guidelines for human papillomavirus DNA test requirements for cervical screening

  Intra-laboratory agreement Inter-laboratory agreement
HPV type Both runs Run 1 only Run 2 only Kappa Both labs Lab 1 only Lab 2 only Kappa
HPV 16 65 1 2 0.974 66 0 2 0.983
HPV 18 14 0 0 1.000 14 0 2 0.931
hrHPV 116 3 3 0.968 118 1 2 0.984