Skip to main content

Table 2 Intra-laboratory reproducibility over time of the REALQUALITY RQ-HPV Screen

From: Clinical validation of REALQUALITY RQ-HPV Screen according to the international guidelines for human papillomavirus DNA test requirements for cervical screening

  First result
Second result Positive Negative Total
Positive 167 2 169
Negative 1 366 367