Skip to main content

Table 4 Summary of LLoQ Results across Genotypes in Plasma and Serum

From: Analytical characteristics and comparative evaluation of Aptima HCV quant Dx assay with the Abbott RealTime HCV assay and Roche COBAS AmpliPrep/COBAS TaqMan HCV quantitative test v2.0

HCV Genotype

Plasma

Serum

LLoQ

LLoQ

LLoQ

LLoQ

(log IU/ml)

(IU/ml)

(log IU/ml)

(IU/ml)

1 (WHO)

0.82

7

0.93

9

2

0.76

6

0.80

6

3

0.80

6

0.67

5

4

0.82

7

0.65

4

5

0.87

7

0.72

5

6

0.79

6

0.99

10