Table 5 Assessment of inter-instrument, inter-operator, inter-lot, inter-day, inter-run, intra-run and total variability of Aptima HIV-1 Quant Dx in various HIV-1 RNA concentrations

From: Analytical characteristics and comparative evaluation of Aptima HIV-1 Quant Dx assay with Ampliprep/COBAS TaqMan HIV-1 test v2.0

   

Inter-instrument

Inter-operator

Inter-lot

Inter-day

Inter-run

Intra-run

Total

Panel

N

Mean

SD

CV (%)

SD

CV (%)

SD

CV (%)

SD

CV (%)

SD

CV (%)

SD

CV (%)

SD

CV (%)

Log

5E1 copies/mL

41a

1.66

0.075

4.52

0.033

1.99

0

0

0

0

0

0

0.119

7.12

0.144

8.67

1E2 copies/mL

74a

1.82

0

0

0

0

0.014

0.77

0

0

0.102

5.6

0.148

8.14

0.18

9.91

1E3 copies/mL

81

2.75

0.035

1.26

0

0

0

0

0

0

0.041

1.5

0.098

3.57

0.112

4.08

1E4 copies/mL

81

3.81

0.011

0.28

0.013

0.35

0

0

0.013

0.35

0.05

1.31

0.039

1.04

0.067

1.77

1E5 copies/mL

81

4.96

0.01

0.21

0

0

0

0

0

0

0.042

0.84

0.051

1.03

0.067

1.35

1E6 copies/mL

78b

6

0.008

0.14

0.001

0.02

0

0

0

0

0.048

0.79

0.027

0.44

0.055

0.92

1E7 copies/mL

81

6.89

0

0

0

0

0.026

0.38

0

0

0.033

0.47

0.046

0.67

0.062

0.9

  1. Note: 1. Variability from some factors may be numerically negative, which can occur if the variability due to those factors is very small. When this occurs, SD = 0 and CV = 0 %
  2. athe 50 and 100 copies/mL sample has lower N because some results were below the limit of quantification
  3. b3 results were invalid and not included in the analysis