Skip to main content

Table 4 Assessment of analytical specificity of Aptima HIV-1 Quant Dx assay by two lots of HIV-1 negative samples

From: Analytical characteristics and comparative evaluation of Aptima HIV-1 Quant Dx assay with Ampliprep/COBAS TaqMan HIV-1 test v2.0

Reagent lot

N

Negative

Specificity

LCI

UCI

lot 1

300

299

99.67

98.14

 99.94

lot 2

300

300

100

98.74

 100.00

Combined

600

599

99.83

 99.06

 99.97