Skip to main content

Table 1 Performance of the Luminex GPP in comparison to the rotavirus reference qRT-PCR assay among all study subjects (top panel) and cases only (bottom panel)

From: Diagnostic performance of the Luminex xTAG gastrointestinal pathogens panel to detect rotavirus in Ghanaian children with and without diarrhoea

  TP, n FP, n FN, n TN, n Sensitivity % (95 % CI) Specificity % (95 % CI) kappa
All study subjects (n = 682) 67 0 9 606 88.2 (78.2–94.1) 100 (99.2–100) 0.93
Cases (n = 341) 44 0 5 292 89.7 (76.9–96.2) 100 (98.4–100) 0.94
  1. TP true positive; FP false positive; FN false negative; TN true negative
\