Skip to main content

Table 4 Concordance between custom and FTD assay

From: Evaluation of custom multiplex real - time RT - PCR in comparison to fast - track diagnostics respiratory 21 pathogens kit for detection of multiple respiratory viruses

Virus

FTD (+) Custom (+)

FTD (+) Custom (-)

Custom (+) FTD (-)

Concordance between custom assay and FTD assay in %

HCoV OC43

7

0

0

100.0

HCoV 229E

3

0

0

100.0

Flu B

12

0

1

99.71

HPIV-1

6

0

0

100.0

HPIV-3

22

1

3

98.87

HPIV-4

5

1

1

99.43

HPIV-2

5

0

0

100.0

HCoV NL63

2

1

0

99.71

HMPV A/B

42

1

0

99.71

RSV A/B

66

1

18

94.66

HBoV

16

0

0

100.0

HPeV

1

0

0

100.0

HRV

65

0

1

99.71

EV

8

2

4

98.31

HCoV HKU1

2

1

0

99.71

HAdV

26

0

1

99.71

Flu A

2

0

0

100.0

Influenza A(H1N1)pdm09

27

0

0

100.0