From: Development of an optimized method for the detection of airborne viruses with real-time PCR analysis
Sample ID | HAdV added GC/plate | HAdV recovered GC/plate | Recovery (%) | Mean recovery (%) |
---|---|---|---|---|
exp1 | 1 × 104 | 0,47 × 104 | 47 | 81 |
exp2 | 1 × 104 | 1,01 × 104 | 101 |  |
exp3 | 1 × 104 | 0,95 × 104 | 95 |  |
exp4 | 1 × 103 | 0,59 × 103 | 59 | 60,3 |
exp5 | 1 × 103 | 0,59 × 103 | 59 |  |
exp6 | 1 × 103 | 0,6 × 103 | 63 |  |
exp7 | 1 × 102 | 0,42 × 102 | 42 | 33,7 |
exp8 | 1 × 102 | 0,26 × 102 | 26 |  |
exp9 | 1 × 102 | 0,33 × 102 | 33 |  |