Table 2 Mean Ct/Cp values and %CV values for the undiluted OptiQual Enterovirus Run Control in different real-time RT-PCR settings

From: Improvement of a real-time RT-PCR assay for the detection of enterovirus RNA

Extraction Method Platform Mean Ct/Cp
(%CV)
   Primer-probe set 1 Primer-probe set 2
QiaAmp 7700 SDS 38.47 ± 3.94
(10.2%)
37.74 ± 1.35
(3.6%)
  LC 480 33.69 ± 2.95
(8.8%)
34.48 ± 1.67
(4.8%)
EasyMag 7700 SDS 38.23 ± 3.88
(10.2%)
36.29 ± 0.91
(2.5%)
  LC 480 32.81 ± 0.25
(0.8%)
32.78 ± 0.20
(0.6%)