Figure 1 | Virology Journal

Figure 1

From: Detection of Middle East respiratory syndrome coronavirus using reverse transcription loop-mediated isothermal amplification (RT-LAMP)

Figure 1

Sensitivity of the MERS-CoV RT-LAMP assay. a) Real-time amplification of MERS-CoV by RT-LAMP. Amplification of serially diluted MERS-CoV RNA was measured in real-time using a Loopamp real-time turbidimeter (LA-320C). The differentiated value at each dilution was calculated automatically, with values > 0.1 within 30 min (red line) considered positive. b, c) Detection of RT-LAMP amplicon by b) precipitation of magnesium pyrophosphate and c) fluorescence under ultra violet light. For fluorescence detection, 1-μL Fluorescence Detection Reagent was added to each reaction mixture. Fluorescent signals were detected under ultraviolet following completion of the amplification reaction. Successful amplification could be detected as green fluorescent light.