Skip to main content

Table 2 SYBR Green PCR assay for TTV using single series of plasmid standards

From: Validation of SYBR Green based quantification assay for the detection of human Torque Teno virus titers from plasma

Plasmid standards

Day

CTMean ± SD (n=3)*

Intra-day precision

Inter-day precision

Intra-day accuracy

Inter-day accuracy

10 × 109

1

4.21 ± 0.02

0.50

0.45

99.6

99.6

2

4.18 ± 0.01

0.39

 

99.6

 

10 × 108

1

7.59 ± 0.05

0.75

0.65

101.3

101.3

2

7.52 ± 0.05

0.78

 

101.2

 

10 × 107

1

11.74 ± 0.05

0.45

1.10

100.8

100.8

2

11.56 ± 0.01

0.14

 

100.8

 

10 × 106

1

15.83 ± 0.04

0.28

1.22

100.4

100.4

2

15.56 ± 0.01

0.08

 

100.4

 

10 × 105

1

19.75 ± 0.05

0.26

1.39

100.6

100.7

2

19.37 ± 0.01

0.10

 

100.8

 

10 × 104

1

24.15 ± 0.03

0.12

1.52

98.1

98.4

2

23.64 ± 0.02

0.10

 

98.6

 

10 × 103

1

28.40 ± 0.03

0.12

1.64

95.3

95.8

2

27.75 ± 0.05

0.19

 

96.4

 

10 × 102

1

33.21 ± 0.20

0.62

1.51

83.8

84.3

2

32.51 ± 0.06

0.21

 

84.8

 

10 × 101

1

35.59 ± 0.39

1.11

0.67

102.1

99.0

 

2

35.25 ± 0.06

0.17

 

95.7

 
  1. * Tenfold dilutions run in triplicates.